Alat Terapi Gerak Jari Tangan untuk Pasien dengan Cedera Plexus Brachialis

Agustine, Lanny, Setiawan, Oscar and Joewono, Andrew (2016) Alat Terapi Gerak Jari Tangan untuk Pasien dengan Cedera Plexus Brachialis. Media Teknika Jurnal Teknologi, 11 (2). pp. 65-75. ISSN 1412-5641

[thumbnail of Alat Terapi Gerak Jari Tangan untuk Pasien dengan Cedera Plexus Brachialis]
Preview
Text (Alat Terapi Gerak Jari Tangan untuk Pasien dengan Cedera Plexus Brachialis)
7-Alat_terapi_gerak_jari_.pdf

Download (889kB) | Preview
[thumbnail of Alat Terapi Gerak Jari Tangan untuk Pasien dengan Cedera Plexus Brachialis_peer_review]
Preview
Text (Alat Terapi Gerak Jari Tangan untuk Pasien dengan Cedera Plexus Brachialis_peer_review)
7-R1&2-Alat_Terapi_Gerak_Jari_.pdf

Download (646kB) | Preview
[thumbnail of Alat Terapi Gerak Jari Tangan untuk Pasien dengan Cedera Plexus Brachialis]
Preview
Text (Alat Terapi Gerak Jari Tangan untuk Pasien dengan Cedera Plexus Brachialis)
7-Alat_terapi_gerak_jari_Hasil Cek Similarity.pdf

Download (3MB) | Preview

Abstract

People with damaged or injured plexus brachialis need immediate therapy after the surgery, as an effort to restore the nerves function, even thou it is hard to achieve. But, continuous treatment may prevent the muscle to become stiff and shrink. So, patient needs assistive device to be used at home for regular self therapy. Physical therapy method was chose to be developed in this research because of lower risk than electrical therapy. This assistive device is designed in two modes, manual and automatic, and it is limited to fingers movement of adult’s right hand. Manual mode is focusing to 1 finger movement, while automatic mode will move the five fingers consecutively. The movement procedure is imitated those given in the clinic, by controlling the speed and angle rotation of a dc servo motor and certain design of lever for each finger. This assistive device works effectively for a certain fingers size. It takes 5.29 sec to finish one cycle in automatic mode, and about 2 sec in manual mode. The degree of finger movement yield by the device is about 20 degree less than finger maximum ROM for MCP and PIP joint of index finger, pinky, and thumb.

Item Type: Article
Uncontrolled Keywords: device, physical, therapy, finger, movement
Subjects: Engineering > Electrical Engineering
Divisions: Journal Publication
Depositing User: F.X. Hadi
Date Deposited: 24 May 2019 08:25
Last Modified: 27 Aug 2019 01:54
URI: http://repository.ukwms.ac.id/id/eprint/18455

Actions (login required)

View Item View Item